UV-3600
High Sensitivity, High Resolution, Ultra-low Stray Light
The UV-3600 UV-VIS-NIR Spectrophotometer is a new additional model in our product lineup of high-end UV-VIS-NIR spectrophotometers.
The UV-3600, a successor model to the UV-3150/3101, is designed for the measurement of liquid samples, while the SolidSpec-3700/3700DUV is mainly designed for the measurement of solid samples.
Features
High Sensitivity - The Only 3 Detector System
Equipped with 3 detectors: A PMT detector (photomultiplier tube) for the ultraviolet and visible regions, and InGaAs and PbS detectors for near infrared region. The InGaAs detector bridges the gap between the PMT - PbS switching wavelength where sensitivity is typically low to ensure high sensitivity over the entire measurement wavelength range. With <0.00003 Abs noise at 1500 nm, the world´s lowest noise level is achieved.
High Resolution - Wide Measurement Range & Ultra-low Stray Light With its high performance double monochromator, ultra-low stray light (0.00005% or less at 340 nm) is achieved at high resolution (highest resolution 0.1 nm). With a measurement wavelength range of 185 - 3300 nm, measurement is possible over a wide range including the ultraviolet, visible and near infrared regions. This allows spectroscopic analysis in a wide variety
of fields.
All-in-one Software
UVProbe software for the UV-3600 includes functions for spectrum measurement, quantitation, kinetics and a
report generator. Multiple security and audit-trail functions ensure the reliability of data.
| Wavelength range | 185nm - 3300nm |
| Spectral Bandwidth | 8 steps in ultraviolet/visible region 0.1, 0.2, 0.5, 1, 2, 3, 5, 8 nm 10 steps in near-infrared region 0.2, 0.5, 1, 2, 3, 5, 8, 12, 20, 32 nm |
| Resolution | 0.1 nm |
| Sampling Pitch | 0.01 - 5 nm |
| Wavelength accuracy | UV/VIS region: ±0.2 nm NIR region: ±0. 8 nm |
| Wavelength repeatability | UV/VIS region: less than ±0.08 nm NIR region: less than ±0.32 nm |
| Stray light | < 0.00008% (220 nm, NaI) |
| < 0.00005% (340 nm, NaNO2) | |
| < 0.0005% (1420 nm, H2O) | |
| < 0.005% (2365 nm, CHCl3) | |
| Photometric range | -6 to 6 Abs |
| Noise | 0.00005 Abs or less (500 nm), 0.00008 Abs or less (900 nm), 0.00003 Abs or less (1500 nm) Slit width 2nm, RMS value at 1 sec. response |
| Photometric System | Double beam |
| Dimensions | 1020(W) x 660(D) x 275(H) mm |